ORP meter with two-point alarm
OE-31E

 2点警報付ORP計 OE-31E

ORP Indication Controller with Two-Point Alarm.

This is an easy-to-use model with a detailed set of settings that can respond to a variety of requests and an error display based on the self-diagnosis function.
ORP計
 
Advantages Applications Specifications Catalogue Relevant  

Advantages

・Automatically judges whether the electrode characteristics are good or bad.

(Out of measurement range, converter error)

・Combined with lead-free electrodes, environmentally friendly measurements are realized.
・Hold function of the last value or arbitrary value.
 

Applications

・In the oxidation and reduction of wastewater from the chemical industry
・For the measurement of oxidation and reduction of various solutions

Specifications

Type OE-31E
Display Digital LED
Measuring range

-2000~+2000mV

Resolution 1mV 
Transmission output

4-20 mADC (max. load-resistance 900 Ω)

Transmission output range: Any setting within the measurement range is possible.

Contact output

Output point: 2 points

Warning contact output :(R1、R2)

Contact form: Relay contact SPDT (1c)

Contact capacitance: 240V AC 3A30V DC 3A (resistive loading)

Holding function

Select the value from the previous value hold or arbitrary value hold (however, in the maintenance mode, the value is the previous hold).

Self-diagnostic functions

Out of measurement range, converter error
Ambient temperature -5~45℃

Relative humidity

20 to 85% (not condensing)

Power

AC100V~240V (50/60Hz) 10VA(max)
Power consumption 10 VA (no load)
Dimensioned 96(W)×96(H)×115(D) mm
Weight About 500 g